Biochemistry Analyzer
Overview
A fully automated POC Auto Dry Biochemistry Analyzer called Mispa Chem Dx, With Just 3 to 4 drops (100ul) of whole blood to test up to 19 General and Special chemistry tests in less than 12 minutes. Mispa Chem Dx works on absorption spectroscopy where a change in absorbance is detected by an absorption spectrum and an algorithm is used to calculate the end concentration of the sample in the reagent dry disc. The reagent disc consists of lyophilized reagent beads, diluent, and calibration details in the QR code. The reagent disc contains an intelligent internal quality control system IQC which continuously monitors the disc function to ensure reaction stability and optimal disc performance, IQC is the engine that drives the Mispa Chem Dx accurately and precisely
Low Sample Consumption
Easy to Use
Maintenance Free
Accurate Result
Disposable reagent disc
Sample volume: 100µl
Testing time: 12 minutes / sample
Temperature: 37°C ± 0.3°C
Resolution: 0.001Abs Absorbance: 0-3.0Abs
Sample type: Heparinized whole blood, serum, plasma (heparin)
QC&Calibration: IQC (Intelligent Quality Control/Calibration)
Work condition: Temperature:10-30°C Humidity: 30%-70%
Testing method: End point, kinetic, fixed time, turbidmetry etc
Light source: 12V/20W, Halogen tungsten lamp with life span over 2500 hours
Power supply: AC 100V-240V, 50-60Hz
Display: Android 7.0 inch 800*480, multi-point capacitive touch screen
Storage: 500,000 results
Printer: Built-in thermal printer Interface: 4 USB ports, 1 LAN port
Dimension: 315*375*475mm. 4.2kg(N.W.), 7kg(G.W.)