Product Features: Better precision, CV < 2.5 % Patented CLE technology enables complete elimination of labile glycohemoglobin ensuring accurate HbA1c results. HbA1c test results in 3.5 minutes with HbA2/HbF flagging. Separation of hemoglobin fractions Complete removal of labile fraction up to 10% No interference of HbF up to 10% No interference of lipemia or icterus, can run on non-fasting patient samples HbA1c reportability presence of HbS, HbC, HbD and HbE Compact bench top model User friendly touch screen operation Sampling from multiple sized primary tubes, including microcapillary collection tubes On board QC functions with Levey-Jennings and statistical calculations Unlimited storage of calibration, QC data and chromatograms No sample preparation required Simple reagent and column replacement Automated maintenance RFID reagent identification and tracking Automatic barcode identification of samples NGSP certified Calibrators and Controls IFCC traceable Complete kit with buffers, calibrator and column Product Application: Hba1c Testing Sampler Handler: 10 Position Rack Loader, positive identification by internal barcode reader Sample Type: Whole blood on EDTA K3 or K2 anticoagulant Haemolysate from EDTA whole blood samples Time for Analysis: Throughput approximately 3.5 minutes between samples Interferences: Complete removal (not separation) of labile up to 10% Sampling System: Sample Aspiration Probe coupled with a high precision, Sample Aspiration Volume: 14.5 µl in whole blood, 670 µl in lysate mode. Pump: 2 high pressure syringe pumps Detector Type: UV-Vis Reagent Traceability: RFID Printer: Built- in thermal printer with possibility to connect to external printer